Industrial Gaussmeter IGM11 Operating Instructions
Chapter 9 Serial Interfaces

9.3.2 Operating Mode SHORT

In this operating mode simple one-character commands are evaluated and replied
depending on the command type. The IGM11 will only reply on request. The com-
mand line must start with <STX> and terminate with <ETX>. The reply will also be
enclosed between <STX> and <ETX>. There is no distinction between upper and
lower case.
Command
<STX>?<ETX>
<STX>R?<ETX>
<STX>R0<ETX>
<STX>R1<ETX>
<STX>R2<ETX>
<STX>R3<ETX>
<STX>O?<ETX>
<STX>N<ETX>
<STX>P<ETX>
<STX>S?<ETX>
Page 47 / 92
Description
Read out current measuring
value
Read out measuring range
Set measuring range to the
most sensitive range (e.g.
10 mT)
Analogous to R0, second
most sensitive range (e.g.
100 mT)
Analogous to R0, third most
sensitive range (e.g.
1000 mT)
Set measuring range to least
sensitive measuring range
(e.g. 4500 mT)
Read out signal outputs
Set probe to zero
Set peak to zero
Read out status
Reply
Current measuring value
s = +|-
n = 0..9
Current measuring range
n = 0..3
None
None
None
None
Status of the signal outputs
One character per output:
0 (0x30h) or 1 (0x31h)
None
None
Current status
s = S|R
S: Stop; R: Run
Reply format
<STX>snnnn.<ETX>
<STX>snnn.n<ETX>
<STX>snn.nn<ETX>
<STX>sn.nnn<ETX>
<STX>b<ETX>
--
--
--
--
<STX>oooo<ETX>
--
--
<STX>s<ETX>