MAGSYS IGM11 Operating Instructions Manual page 78

Hide thumbs Also See for IGM11:
Table of Contents

Advertisement

9.8.7.10 :DGIO:SET:IN
Description:
Mode:
Parameter:
Response:
*RST
Example:
9.8.7.11 :DGIO:SET:OUT
Description:
Mode:
Parameter:
Response:
*RST
Example:
Page 78 / 92
{x}
Set or query the logic level of a digital input.
x is for inputs 1 to 3.
Command or query
{
}
0|1
0 is for normal function (current flow sets the input)
1 is for inverted function
like parameter
Value:
not relevant
send
:DGIO:SET:IN2:LOG 1<LF>
send
:DGIO:SET:IN2:LOG?<LF>
receive
1<CR><LF>
Set or query the function of each output.
x is for outputs 1 to 4.
Command or query
{
NO|IFHI|IFLO|IFIN|IFOU|IFOV|IFRN
Parameter
NO
IFHI
IFLO
IFIN
IFOU
IFOV
IFRN
like parameter
Value:
not relevant
send
:DGIO:SET:OUT2:FUNC IFOV<LF>
send
:DGIO:SET:OUT2:FUNC?<LF>
receive
IFOV<CR><LF>
Gaussmeter IGM11 Operating Instructions
:LOG[?]
:FUNC[?]
{x}
Function
No function assigned to this output
This output is set if the measuring value is geater
than the set upper limit value.
This output is set if the measuring value is less
than the set lower limit value.
This output is set if the measuring value is within
the range of upper and lower limit values.
This output is set if the measuring value exceeds
the range of lower and upper limit values.
This output is set if the measuring range has been
exceeded.
This output is set if the device is ready to measure.