Industrial Gaussmeter IGM11 Operating Instructions
Chapter 6 Set-up Menu

6.3.5 Settings of the Digital Inputs

Correspondingly, these settings hold for all three inputs. If two inputs are configured
with the same function, both functions are evaluated as OR-operation.
Page 37 / 92
Settings of digital input 1
No function
NONE
Hold
HOLD
Reset of the peak
NULL PEAK
value
Probe null compen-
NULL SET
sation
Logic level
24V
Logic level
0V
This input is not used
The measurement is on hold
if the input is set
The peak value is set to null
(edge).
The null compensation will be
started (edge).
This function will be started
with rising edge or active sig-
nal.
This function will be started
with falling edge or no signal.