Memory Functions - MAGSYS IGM11 Operating Instructions Manual

Hide thumbs Also See for IGM11:
Table of Contents

Advertisement

9.8.5 Memory Functions

9.8.5.1 :PAR:SAVE
Description:
Mode:
Parameter:
Response:
*RST
Example:
9.8.5.2 :PAR:SAVW?
Description:
Mode:
Parameter:
Response:
*RST
Example:
9.8.5.3 :PAR:DEFA
Description:
Mode:
Parameter:
Response:
*RST
Example:
Page 72 / 92
Save parameters, units and ranges in non-volatile memory.
Command
none
none
Value:
not relevant
send
:PAR:SAVE<LF>
Save parameters, units and ranges in non-volatile memory. After that,
the parameter status will be returned.
Query
none
: Storage error
0|1 1
Value:
not relevant
send
:PAR:SAVW<LF>
receive
0<CR><LF>
Parameters, units and ranges are reset to default.
Command
none
none
Value:
not relevant
send
:PAR:DEFA<LF>
Gaussmeter IGM11 Operating Instructions
Chapter 9 Serial Interfaces

Advertisement

Table of Contents
loading

Table of Contents