MAGSYS IGM11 Operating Instructions Manual page 61

Hide thumbs Also See for IGM11:
Table of Contents

Advertisement

Industrial Gaussmeter IGM11 Operating Instructions
Chapter 9 Serial Interfaces
9.8.1.4 *ESR[?]
Description:
Mode:
Parameter:
Response:
*RST
Example:
9.8.1.5 *IDN?
Description:
Mode:
Parameter:
Response:
*RST
Example:
9.8.1.6 *OPC?
Description:
Mode:
Parameter:
Response:
*RST
Example:
Page 61 / 92
Standard Event Status Enable Query
Reset or query the standard event release register. The gaussmeter
returns the bit pattern of the register as a decimal value.
Command and query
{<Value>} within the range of 0..255
Standard event release register
Value:
not relevant
send
send
receive
Identification Query
Query the gaussmeter identification character string. The gaussmeter
returns the following identification text:
MAGSYS-MAGNET-SYSTEME,IGM11,1,xxxxxx,C
Query
none
Identification character string
Value:
not relevant
send
receive
MAGSYS-MAGNET-SYSTEME,IGM11,1,011111,C<CR><LF>
Operation Complete Query
Return of "1" to the output buffer, after the command has been exe-
cuted. The command *OPC? can be placed at the end of a command
line in order to synchronize the controller with the command pro-
cessing. The command itself has no function.
Query
none
1
Value:
not relevant
send
receive
*ESR 22<LF>
*ESR?<LF>
160<CR><LF>
*IDN?<LF>
*OPC?<LF>
1<CR><LF>

Advertisement

Table of Contents
loading

Table of Contents