Chapter 9 Serial Interfaces
9.8.7.7 :DGIO:IN1? .. :DGIO:IN3?
Description:
Mode:
Parameter:
Response:
*RST
Example:
9.8.7.8 :DGIO:IN?
Description:
Mode:
Parameter:
Response:
*RST
Example:
9.8.7.9 :DGIO:SET:IN
Description:
Mode:
Parameter:
Response:
*RST
Example:
Page 77 / 92
Query the digital input state.
Query
none
if the input is current-carrying
0|1 1:
Value:
not relevant
send
receive
Query the states of all digital process inputs.
Query
none
1 if the input is current-carrying
x;y;z x=
x is for input 1; y for input 2 etc.
Value:
not relevant
send
receive
{x}
Set or query the function of a digital input.
x is for inputs 1 to 3.
Command or query
{
NO|HOLD|NLPK|NULL
Parameter
NO
HOLD
NLPK
NULL
like parameter
Value:
not relevant
send
send
receive
:DGIO:IN1?<LF>
1<CR><LF>
:DGIO:IN?<LF>
1;0;1<CR><LF>
:FUNC[?]
}
Function
No function assigned to this input
Setting this input will stop the measurement
Setting this input will reset the peak values (with
rising edge)
Setting this input will zero compensate the input
(with edge)
:DGIO:SET:IN2:FUNC NLPK<LF>
:DGIO:SET:IN2:FUNC?<LF>
NLPK<CR><LF>