Power Supply Connection - MAGSYS IGM11 Operating Instructions Manual

Hide thumbs Also See for IGM11:
Table of Contents

Advertisement

Power Supply Connection

Via connector J6 the device is supplied with DC power.
The operating voltage must be between 12 V and 24 V. The power consumption is
120 mA maximum at a voltage of 12 V.
The connector is protected against wrong polarity connection.
In a switch cabinet, the power supply system should be fused.
The common potentials of the input and output signals can easily be connected to
supply voltage via the double connector plug.
Page 40 / 92
The ground connection (GND) of the power supply is galvanically
connected with the measuring probe and the serial interfaces.
Note
Please pay attention that this connector is connected with the earth
conductor inside the switch cabinet. Please also ensure that neither
the power supply nor the measuring probes nor the interface wiring
can be connected with dangerously high voltages.