9.6.4.5 Definition of the Register Bits
Definition of the bits in the measurement event register
Bit
Definition of the bits in the error event register
Bit
Definition of the bits in the standard event register
Bit
Definition of the bits in the sum register
Bit
0
3
5
6
Page 56 / 92
Decimal
Description
Value
0
1
Overflow
1
2
Data available
Decimal
Description
Value
1
2
General
Calibration Error
6
64
Internal
Calibration Error
7
128
Probe
Calibration Error
Decimal
Description
Value
0
1
Command executed
(OPC, operation complete)
3
8
Device error
(DDE, device-dependend
error)
5
32
Command error
(CME, command error)
7
128
Ready for use
(PON, power-on)
Decimal
Description
Value
1
Measurement
event
8
Error event
32
Standard event
64
Service request
Gaussmeter IGM11 Operating Instructions
Chapter 9 Serial Interfaces
Definition
Range overflow during measurement
Measurement terminated. Data available.
Definition
Is set if the calibration data of the measuring
probe could not be read or if the internal calibra-
tion data are not consistent.
Is set if the internal calibration data are not con-
sistent.
Is set if the calibration data of the measuring
probe are not consistent.
Definition
Is set at the end of the SCPI command processing
An error has occured.
An error has occcured during the SCPI command
processing.
Is set when the device is ready for use.
Definition
One or several bits are set in the measurement
event register and activated in the release register.
One or several bits are set in the error event register
and activated in the release register.
One or several bits are set in the standard event
register and activated in the release register
One or several bits are set in the sum register and
activated in the release register.