Power Supply; Probe Connection; Ports Bottom - MAGSYS IGM11 Operating Instructions Manual

Hide thumbs Also See for IGM11:
Table of Contents

Advertisement

Industrial Gaussmeter IGM11 Operating Instructions
Chapter 4 Control Elements and Connections
For add-ons
Do not use

Power Supply

The gaussmeter can be operated at a voltage of 12 V to 24 V DC. The current con-
sumption is approx. 120 mA at 12 V or approx. 85 mA at 24 V. An electrical sepa-
ration does not take place. Please notice that the shielding of the probes (if so, the
metal housing) will be connected with the earth conductor of your control via this
port.

Probe Connection

The measuring probe is plugged into the probe port (M8-socket IEC 61076-2-104)
on the top of the device.
Page 19 / 92
Figure 8 Ports Bottom
Interface
Port for PLC sig-
nals
Note
Only use measuring probes which are approved by the manufac-
turer to operate with this device.
The probe plugs must not be connected to the electric potential, the
Note
protective conductor or the plug shell. If you measure near current
conducting parts, make sure that there is sufficient distance and a
sufficiently good insulation.
Power Supply
LAN
Port for power sup-
RJ45 port for the LAN
ply for the device
connection
and the PLC signals

Advertisement

Table of Contents
loading

Table of Contents