Industrial Gaussmeter IGM11 Operating Instructions
Chapter 9 Serial Interfaces
register, a decimal value is transmitted that corresponds to the sum of the binary place
values of all bits set in this register.
9.6.4.3 Sum Register
In the sum register the results of the event registers are summarized after the masking
in the release register and assigned to single bits. The deletion of the event in the
event registers also sets back the respective sum bit in the status register. Via the re-
lease register of the sum register the single sum bits can be re-merged to one whole
message service query.
9.6.4.4 Overview Status Model
Measurement
Event Register
Release Register
0
Overload
1
Data ready
Not used
Not used
Not used
Not used
Not used
Not used
STAT:MEAS:EVEN?
STAT:MEAS:ENAB < Value >
STAT:MEAS:ENAB?
Error Data
Event Register
Release Register
Not used
1
Calibration Error
Not used
Not used
Not used
Not used
6
Internal Cal. Error
7
Probe Cal. Error
STAT:QUES:EVEN?
STAT:QUES:ENAB < Value >
STAT:QUES:ENAB?
Standard
Event Register
Release Register
0
Operation ended
Not used
Not used
3
Internal Error
Not used
5
SCPI-Error
Not used
7
On
*ESR?
*ESE < Value >
*ESE?
Page 55 / 92
OR
Sum Register
0
Meas. Event
Not used
OR
Not used
3
Error Data
Not used
5
Standard Event
6
Service-Request
Not used
*STB?
OR
Release Register
OR
*SRE < Value >
*SRE?