Power Supply Connection; Replacing A Fuse - GE Druck ADTS 542F Safety And Installation Manual

Air data test set
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
1.4 Power Supply
Single phase

1.5 Power Supply Connection

Do not apply electrical power outside the values stated.
Make sure that the power supply is off before connecting the power cable.
The ADTS must be connected to the correct electrical power supply as stated, adjacent to the
power connector.
RISK OF ELECTRIC SHOCK
electrocution if there is not a protective earth in place. The ADTS must, at
all times, be connected to the supply protective earth/ground.
The power supply cable and connector must be correctly rated for the
power supply. Only use power cables supplied by GE specifically for the
ADTS.
European Color
Green/Yellow
1.6 Fuses
The two High Breaking Capacity fuses (8) (see Figure 3.1), located in the holders and mounted
on the front panel, protect the unit. The fuses are connected in the live and neutral supply
circuit and are rated at: T5AH 250V.

1.7 Replacing a Fuse

1.
Disconnect the electrical power supply.
2.
Unscrew the fuse holder cap from the front panel and remove the fuse.
3.
Fit a new fuse of the correct size and rating into the fuse holder cap.
4.
Secure the fuse holder cap in the front panel.
© 2014 General Electric Company – All rights reserved.
110/230 Vac, 50/60Hz
115 Vac, 400Hz
US Color
Brown
Blue
200 VA Max - ADTS542F/ADTS552F
300 VA Max - ADTS553F/ADTS554F
There is a risk of electric shock/
Black
White
Green
Protective Earth (Ground)
Power Supply
Function
Live
Neutral
ADTS Safety and Installation Guide-English | 5