Hach D3433B8.99 User Manual page 17

Contacting conductivity sensors
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Warning
The device is not
calibrated.
Replace a sensor.
Calibration is in
progress...
Outputs on hold
Linear temperature
compensation is out of
range.
Temperature
compensation table is
out of range.
6.6 Event list
The diagnostic bar shows current activities such as configuration changes, alarms, warning
conditions, etc. A list of possible events is shown in
event log, which can be downloaded from the controller. Refer to the controller documentation for
data retrieval options.
Event
Calibration ready
The calibration is OK.
The time has expired.
The calibration failed.
The calibration is high.
K is out of range.
The reading is unstable.
Change in configuration float value
Change in configuration text value
Change in configurationint value
Change in configuration
Power is on.
ADC failure
Flash erase
Temperature
The sample calibration was started.
Table 6 Warning list (continued)
Description
The sensor has not been calibrated.
The Sensor days counter is more
than the interval selected for sensor
replacement. Refer to
Configure the
sensor
on page 8.
A calibration was started but not
completed.
During calibration, the outputs were
set to hold for a selected time.
The user-defined linear temperature
compensation is out of range.
The user-defined temperature
compensation table is out of range.
Table 7 Event list
Description
The sensor is ready for calibration.
The current calibration is good.
The stabilization time during calibration expired.
The calibration failed.
The calibration value is above the upper limit.
The cell constant K is out of range for the current calibration.
The reading during calibration was unstable.
The configuration was changed—floating point type.
The configuration was changed—text type.
The configuration was changed—integer value type.
The configuration was reset to the default options.
The power was turned on.
The analog to digital conversion failed (hardware failure).
The flash memory was erased.
The recorded temperature is too high or too low (-20 to 200 °C).
Start of calibration for conductivity
Resolution
Calibrate the sensor.
Replace the sensor. Reset the Sensor days
counter on the Diagnostics/Test > Reset
menu (or Diagnostics/Test > Counter menu.
Return to calibration.
The outputs will become active after the
selected time period. As an alternative,
remove and then supply power to the
controller.
The value must be between 0 and 4%/°C; 0 to
200 °C.
The temperature is above or below the
temperature range defined by the table.
Table
7. Previous events are recorded in the
English 17

Advertisement

Table of Contents
loading

This manual is also suitable for:

D3455a6.993422b6aD3444b8.99

Table of Contents