Expert → Diagnostics → Simulation → Switch status
Switch output simulation (→ 169) = On
Select the status of the status output for the simulation.
• Open
• Closed
The switch status assumes the value defined in this parameter. This helps to check correct
operation of connected control units.
Expert → Diagnostics → Simulation → Sim. alarm
Switch the device alarm on and off.
• Off
• On
When selecting the On option, the device generates an alarm. This helps to check the
correct output behavior of the device in the case of an alarm.
An active simulation is indicated by the C484 Simulation failure mode diagnostic
message.
Expert → Diagnostics → Simulation → Sim. diag. event
Select a diagnostic event for the simulation process that is activated.
When operated via the local display, the selection list can be filtered according to the event
categories (Diagnostic event category parameter).