Operating menu
Switch-off delay
Navigation
Prerequisite
Description
User entry
Factory setting
Failure mode
Navigation
Description
Selection
Factory setting
Switch status
Navigation
Description
Invert output signal
Navigation
Description
Selection
Factory setting
158
Setup → Advanced setup → Switch output → Switch-off delay
• Switch output function (→ 154) = Limit
• Assign limit (→ 155) ≠ Off
Define switch-off delay.
0.0 to 100.0 s
0.0 s
Setup → Advanced setup → Switch output → Failure mode
Define output behavior in alarm condition.
• Actual status
• Open
• Closed
Open
Setup → Advanced setup → Switch output → Switch status
Displays the current state of the switch output.
Setup → Advanced setup → Switch output → Invert outp.sig.
Specify whether the output signal is to be inverted.
• No
• Yes
No
Levelflex FMP55 HART
Endress+Hauser
Need help?