Levelflex FMP53 PROFIBUS PA
Define access code
Navigation
Description
User entry
Additional information
Device reset
Navigation
Description
Selection
Additional information
Endress+Hauser
"Administration" submenu
Navigation
Setup → Advanced setup → Administration → Def. access code
Define release code for write access to parameters.
0 to 9 999
If the factory setting is not changed or if "0" is entered, the parameters are not write-
protected and the device configuration data can therefore always be modified. The
user is logged on in the "Maintenance" role.
The write protection affects all parameters marked with the
document. On the local display, the
the parameter is write-protected.
Once the access code has been defined, write-protected parameters can only be
modified if the access code is entered in the Enter access code parameter
(→ 112).
Please contact your Endress+Hauser Sales Center if you lose the access code.
If operating via the local display: the new access code is only valid once it has been
confirmed in the Confirm access code parameter (→ 153).
Setup → Advanced setup → Administration → Device reset
Reset the device configuration - either entirely or in part - to a defined state.
• Cancel
• To factory defaults
• To delivery settings
• Of customer settings
• To transducer defaults
• Restart device
Meaning of the options
• Cancel
No action
• To factory defaults
All parameters are reset to the order-code specific factory setting.
• To delivery settings
All parameters are reset to the delivery setting. The delivery setting may differ from the
factory default if customer specific settings have been ordered.
This option is only visible if customer specific settings have been ordered.
Setup → Advanced setup → Administration