Level; Medium Type; Medium Property - Endress+Hauser Levelflex FMP53 Operating Instructions Manual

Profibus pa guided wave radar
Hide thumbs Also See for Levelflex FMP53:
Table of Contents

Advertisement

Levelflex FMP53 PROFIBUS PA

Medium type

Navigation
Description
User interface
Factory setting
Additional information

Medium property

Navigation
Prerequisite
Description
Selection
Factory setting
Endress+Hauser
"Level" submenu
Navigation
 Setup → Advanced setup → Level
 Setup → Advanced setup → Level → Medium type
Specify type of medium.
• Liquid
• Solid
FMP50, FMP51, FMP52, FMP53, FMP54, FMP55: Liquid
This parameter determines the value of several other parameters and strongly
influences the complete signal evaluation. Therefore, it is strongly recommended not
to change the factory setting.
 Setup → Advanced setup → Level → Medium property
EOP level evaluation ≠ Fix DC
Specify the dielectric constant ε
• Unknown
• DC 1.4 ... 1.6
• DC 1.6 ... 1.9
• DC 1.9 ... 2.5
• DC 2.5 ... 4
• DC 4 ... 7
• DC 7 ... 15
• DC > 15
Depends on the Medium type (→  113) and Medium group (→  102) parameters.
of the medium.
r