6.2.37 System Failure 63
Cause(s)
Control board has
incompatible software
version.
6.2.38 System Failure 70
Cause(s)
Unexpected SBC software
problem
6.2.39 System Failure 71
Cause(s)
Unexpected SBC software
problem
6.2.40 System Failure 81
Cause(s)
Unexpected SBC software
problem
6.2.41 System Failure 82
Cause(s)
Unexpected SBC software
problem
Service Manual
Action(s)
Replace control board software with latest version
Action(s)
Power cycle the device. If issue persists replace SBC.
Action(s)
Power cycle the device. If issue persists replace SBC.
Action(s)
Power cycle the device. If issue persists replace SBC.
Action(s)
Power cycle the device. If issue persists replace SBC.
Chapter 6: Alarms and System Failure Codes
System Failure Codes
6-25