Incubator Heater Replacement; End Of The Bed; Removing The Incubator Heater Closest To The Control Panel End Of The Bed - GE Giraffe OmniBed Carestation CS1 Service Manual

Table of Contents

Advertisement

7.9.3 Incubator Heater Replacement

Refer to
"Figure 7-47 Heat Sink and
Incubator heater
Motor shaft
Before disconnecting the power, raise the bed to a comfortable working height.
7.9.3.1 Removing the Incubator Heater Closest to the Humidifier End of the Bed
1.
Remove the chassis cover. (Refer to "Removing the Chassis Cover with the Storage Drawer in Place" on
page 7-45.)
2.
Remove the screw (4 mm hex key) in the heatsink directly in front of the heater cartridge.
3.
Disconnect the heater's electrical connector.
4.
Remove the retaining clip and slide the heater out of the heat sink.

7.9.3.2 Removing the Incubator Heater Closest to the Control Panel End of the Bed

1.
Remove the rotating bed, translation deck, tilt platform, and pan.
2.
Remove the fan, fan hub, and fan seal from the top of the fan motor shaft.
3.
Remove the chassis cover. (Refer to "Removing the Chassis Cover with the Storage Drawer in Place" on
page 7-45.)
4.
Disconnect the heater's electrical connector, the thermostat connections and the heat sink sensor
connector.
5.
Remove the upper shroud.
Service Manual
Fan", and
"Figure 8-14 Upper
Fan hub
Chassis
Slot
FIGURE 7-47. Heat Sink and Fan
Chapter 7: Repair Procedures
Chassis".
Fan
Heat sink
Heat sink gasket
Canopy finger switch
Flexible membrane
Blue connector
Lower Device Repairs
Incubator heater
(not shown)
Fan seal
7-47