Figure 7-8. Using North Seal To Help Position Washers; Figure 7-9. Seal Cross-Section - GE Giraffe OmniBed Carestation CS1 Service Manual

Table of Contents

Advertisement

Note:
The cutouts in the north seal may be used to help with this positioning of these washers as shown in
the following figure:
c.
Thread the set screw into the threaded hole in the guide rail near the "E" or "W" inscription. Refer to
Figure
d.
At the middle position of the long canopy sides, install the longer screw, lockwasher, spacer
washer, gasket, door support, and smooth nut, as shown in
e.
Slide the seals on both sides, for alignment purposes. Lift the seal assembly to the upper limit of
the canopy holes. Refer to
f.
Using a 10 mm open-end wrench, tighten all nuts and screws until they feel snug.
g.
Remove seal and verify that the lock washers are compressed flat. If necessary, tighten with
10 mm open-end wrench and 2.5 mm hex key.
Service Manual

FIGURE 7-8. Using North Seal to Help Position Washers

7-7. The socket of the set screw should be on the outside of the canopy.
INSIDE OF CANOPY
Canopy edge
Figure
Caution: Do not over-tighten the hardware.

FIGURE 7-9. Seal Cross-section

7-9.
Chapter 7: Repair Procedures
Canopy Removal for Service
Long lip of seal
Larger lobe
NOTE: The canopy
is upside-down.
Lift assembly to upper limit of canopy holes.
OUTSIDE OF CANOPY
Figure
7-7.
7-7