6.1.6 Bed Height Pedals Disabled
Cause or Circumstances
The Bed Height Pedals Disabled
alarm activates if the bed raise
or lower pedals have been
disabled on either the Service
screen or the Setup screen, and
one of the pedals is pressed.
6.1.7 Bed Raise Pedal Pressed
Cause or Circumstances
The Raise Pedal Pressed alarm
activates if the pedals are
pressed while turning on the
device.
6.1.8 Bed Lower Pedal Pressed
Cause or Circumstances
The Lower Pedal Pressed alarm
activates if the pedals are
pressed while turning on the
device.
Service Manual
Failure
Bed pedal is disabled in service
Mode.
Bed pedal is disabled in Setup
screen.
Failure
Pedals are pressed while
turning on the device.
Failure
Pedals are pressed while
turning on the device.
Chapter 6: Alarms and System Failure Codes
Action
Release the pedal.
If the canopy pedal is disabled in
the service screen and you wish to
use it, contact authorized service
personnel to reactivate it on the
service screen.
Action
Release the pedal.
Action
Release the pedal.
Alarms