GE Giraffe OmniBed Carestation CS1 Service Manual page 298

Table of Contents

Advertisement

Appendix B: Specifications
Installation
Non-Permanent Installation of Equipment
System Performance
Expected Service Life
Patient Temperature Measurement
Accuracy*
Temperature Control Accuracy*
Variability
Warm-up Time
Air Velocity
CO
Level
2
Sound Level (within patient
compartment)
Sound Level for Alarms
USB Port Output
Note:
The above items noted with an asterisk (*) indicate essential performance.
Patient Contact Surfaces
Mattress
Skin Temperature Probe
B-2
Approximately 7 years.
The device is designed to last at least 7 years in normal use when operated,
maintained, and serviced in accordance with the instructions provided in the
provided manuals. With proper maintenance and repairs, the service life can be
extended as long as service parts are available.
± 0.3°C @ 30°C to 42°C
± 1.0°C
± 0.5°C
< 50 min.
Unit with low noise fan: < 30 min.
≤ 10 cm/s
0.3%
Unit with low noise fan: ≤ 41 dBA
Unit without low noise fan: < 50 dBA
63 dBA (High Priority)
60 dBA (Medium Priority)
0.5 A, 5 VDC
Accuracy of patient temperature
measurement system within
range of temperature
measurement.
Control temperature vs average
incubator temperature with level
bed in Manual Mode.
Incubator temperature vs
average incubator temperature.
Time to reach 39°C control temp
from cold start in 25°C 50% RH
room ambient.
Measured in accordance with
IEC 60601-2-19.
In Whisper Quiet™ mode velocity
measured 10 cm above the
center of the mattress, closed
bed.
Maximum CO
level measured per
2
IEC 60601-2-19.
In Whisper Quiet™ mode closed
bed sound level measured 10 cm
above the center of the mattress.
Sound pressure level is measured
in a free field over a reflecting
plane in accordance with
IEC 60601-1-8.
Service Manual