Alarm Test; Diagnostic Readings; Status; Figure 5-6. Error Screen - GE Giraffe OmniBed Carestation CS1 Service Manual

Table of Contents

Advertisement

The Error screen lists a chronological log of the last 32 system errors that occurred with error number and
date.

5.8 Alarm Test

Push and Hold the alarm test button on the Setup screen to test the functionality of the alarm light
indicators, buzzer, and green power status LED on alarm board. It also tests the speakers on the Single
Board Computer and device. See "Setup" on page 4-4.
Note:
Speakers may not be installed.

5.9 Diagnostic Readings

The Diagnostics Readings screen provide a variety of status as well as the ability to test DAC voltage.

5.9.1 Status

A number of status items are available on the Diagnostic readings screen.
Service Manual

FIGURE 5-6. Error Screen

Chapter 5: Troubleshooting
Alarm Test
5-25