GE Giraffe OmniBed Carestation CS1 Service Manual page 98

Table of Contents

Advertisement

Chapter 5: Troubleshooting
Servo Controlled Oxygen
Item
Oxygen0
Oxygen1
Sensor0
Sensor1
Temp
Humidity
5-16
FIGURE 5-1. Servo 02
Description
Oxygen reading in percent from first sensor cell:
Oxygen reading in percent from second sensor cell:
Voltage output in millivolts from first sensor cell.
Voltage output in millivolts from second sensor cell.
Note:
The following information applies to the oxygen0, oxygen1,
sensor0, and sensor1 values:
5 - 90 mv = 21 - 100% O2.
Temperature in degrees Celsius read from a sensor inside the sensor
housing located beneath the bed. Refer to temperature to resistance
curve in the Tips section (5.4.4).
Relative humidity in percent read from a sensor inside the sensor
housing located beneath the bed.
Service Manual