Replacing Rollers And Tension Springs; Separating The Lift Rail From The Belt Channel; Figure 7-26. Rollers And Tensioning Spring - GE Giraffe OmniBed Carestation CS1 Service Manual

Table of Contents

Advertisement

Chapter 7: Repair Procedures
Right Rail Assembly Repair Procedures

7.4.3.2 Replacing Rollers and Tension Springs

Refer to
"Figure 7-26 Rollers and Tensioning
Roller mounting screw
1.
The rollers can be removed one at a time by removing the screw (4 mm hex key) at their hub and pulling
off the roller and bearing. When reinstalling, torque the screws to 45 in. lbs (reference).
2.
To replace the roller tensioning spring, first use needle nose pliers to pull the tension spring off the roll
pin at the end of the lift rail. Then take off the 2 center rollers and slide the tensioning plate out of the lift
rail. Remove the old spring.
3.
Reinstall the 2 center rollers. Use a hex key in the center access slot to lift the tensioning plate up so
that you can screw the rollers into the holes in the plate. Torque rollers screws to 45 inch lbs (reference).
4.
Using pliers, pull the tensioning spring up onto the roll pin at the end of the rail.

7.4.3.3 Separating the Lift Rail from the Belt Channel

Refer to
"Figure 7-25 Right Rail
1.
If you can access the screw head through the access hole in the lift rail go to step 2. If you can not
access the screw, push the lift rail and the belt channel together and lift the belt block up past the roll
pin so you can see the screw head through the access hole in the lift rail.
7-24
Spring".
STEP 3
Bearing
Center rollers
Tensioning plate
Roller tensioning spring

FIGURE 7-26. Rollers and Tensioning Spring

Disassembly".
Lift rail
Service Manual