Starting the VVT
The following procedures explain how to start the Ventilator Verification Test
(VVT).
To start the VVT procedures:
CAUTION
The Alarm Sound Test
The following procedures explain how to perform the Alarm Sound Test.
O2 supply
O2 hose
1.
Connect a breathing circuit to the iVent
2.
Make sure the ventilator is plugged into a working AC supply and switch
on the ventilator's power.
3.
Select a patient weight to enable the ventilator to enter standby mode.
Prior to performing the VVT, the unit must be in a "warm" condition.
Connect the breathing circuit to a test lung and operate the ventilator
for at least 15 minutes.
4.
Proceed to the Main Menu from the Main screen.
5.
From the Main menu, select Maintenance.
6.
A caution pop-up window appears. Select Yes to proceed.
7.
Dial the Control Knob to select Ventilator Verification Test, and press.
8.
The Ventilator Verification pop-up window appears (Figure 158).
Figure 158: The Ventilator Verification Pop-Up Window
9.
Select Start and press the Control Knob.
181
7 Care, Maintenance and Tests
201.
TM
Need help?
Do you have a question about the iVent 201 Series and is the answer not in the manual?
Questions and answers