Aoutput[:Normal]:Preset; Aoutput[:Normal]:Rate<X - Teledyne Everywhereyoulook T3PM1100 User Manual

Digital power meter
Hide thumbs Also See for Everywhereyoulook T3PM1100:
Table of Contents

Advertisement

Example
Note

:AOUTput[:NORMal]:PRESet

Description
Syntax
Parameter
Example
:AOUTput[:NORMal]:RATE<x>
Description
Syntax
Query Syntax
Parameter/
Return parameter
130
:AOUTPUT:NORMAL:MODE1 FIXED
:AOUTPUT:NORMAL:MODE1?
->:AOUTPUT:NORMAL:MODE1 FIXED
• FIXed = Fixed range mode (default value)
Outputs +5 V when the rated value of each
measurement function is received.
• MANual = Manual range mode
The displayed values of the measurement function
when +5 V and –5 V are output as D/A output can
be set to any values of your choice. This enables
the D/A output to be expanded or reduced for each
channel (D/A zoom).
• COMPare = Comparator mode
By comparing with the comparator limits, this
instrument outputs +5 V, 0 V, or –5 V.
Sets the D/A output items to their default values.
:AOUTput[:NORMal]:PRESet {NORMal|INTEGrate}
NORMal
Sets CH1:U, CH2:I, CH3:P, CH4:FU
INTEGrate
Sets CH1:P, CH2:WH, CH3:AH, CH4:FU
:AOUTPUT:NORAML:PRESET NORMAL
Sets or returns the maximum and minimum
values for when the D/A output is in manual
range mode. (Remote Control Only)
:AOUTput[:NORMal]:RATE<x> {<NRf>,<NRf>}
:AOUTput[:NORMal]:RATE<x>?
<x>
1 to 4 (channel)
<NRf>
-9.999E+12~9.999E+12
T3PM1100 User Manual
Set
Set
Query

Advertisement

Table of Contents
loading

Table of Contents