Reset - Emerson Rosemount OCX8800A Manual

Oxygen and combustibles transmitter
Table of Contents

Advertisement

Configuration and start-up
December 2022
3.3.3 
Set test gas values with the local operator interface
(LOI)
Procedure
1. Use the "Z" pattern to enter the LOI menu tree.
2. From the SYSTEM menu, select Calib Setup.
3. From Calib Setup, select O2 High Gas %. Enter the percent O
4. Press Down, and the next selection will be O2 Low Gas %. Enter the percent
5. Press Down several times to display COe Test Gas. Enter the CO
3.4 

Reset

Whenever you correct an equipment alarm or fault condition, the transmitter
reverts to normal operation or continues to indicate an alarm status condition.
If the equipment does not revert to normal operation when a fault condition is
cleared, or if instructed to do so in Troubleshoot, use the following procedure to
reset the transmitter.
3.4.1 
Reset with the local operator interface (LOI)
Procedure
1. Use the Z pattern to enter the LOI menu tree.
2. Select the SYSTEM submenu.
3. From the SYSTEM submenu, select the Status submenu.
4. From the Status submenu, select Reset Device.
The transmitter resets, and the LOI reverts to the normal operation display.
3.4.2 
Reset with Field Communicator
Remove the transmitter from the process loop and recycle power.