Emerson Rosemount OCX8800A Manual page 100

Oxygen and combustibles transmitter
Table of Contents

Advertisement

Calibration
December 2022
6.4.4 
Calibrate O
Communicator: HART
To perform a manual O
AMS, use the following procedure.
Note
To select a menu item, use either Up or Down to scroll to the menu item and
press Right or use the number keypad to select the menu item number.
To return to a preceding menu, press the Left.
Procedure
1. Select CALIBRATION from the DIAG/SERVICE menu.
2. Select PERFORM CAL from the CALIBRATION menu.
3. Select CAL METHODS from the PERFORM CAL menu.
4. From the CAL METHODS menu, select the type of calibration desired: O2 and
5. Remove the transmitter from any automatic control loops to avoid a
6. Set the main Calibration screen to the following values. Press OK to
7. From the SELECT ACTION screen, select START/NEXT CALSTEP to continue
8. When the Calibration Status is at the AppO2Low step, switch on O
100
and COe manually using the Field
2
and COe calibration using the Field Communicator or
2
COe Calibration.
In the first Calibration screen, a Loop should be removed from automatic
control warning appears.
potentially dangerous operating condition and press OK.
continue.
OCX: TAG NAME
STATUS: Idle
TIME REMAIN: 0s
O2: 0.4%, 85.95mV
COe: 0.20 ppm, 0.00 mV
OK/NEXT to Select
ABORT/CANCEL to Exit
calibration, select ABORT CAL to abort calibration, or select EXIT CAL to exit
calibration. Select one from the list and press ENTER.
OCX: TAG NAME
SELECT ACTION
1. START CAL/NEXT CALSTEP
2. ABORT CAL
3. EXIT CAL
Verify the O
concentration measured matches the O2 LOW GAS parameter
2
in Setup. Press OK when ready.
®
Manual
00809-0500-4880
low gas.
2
Emerson.com/Rosemount

Advertisement

Table of Contents
loading

Table of Contents