Ordering Information - Emerson Rosemount OCX8800A Manual

Oxygen and combustibles transmitter
Table of Contents

Advertisement

Manual
00809-0500-4880
1.3.18 
Electrical noise
EN 61326-01, Class A
1.3.19 
Line voltage
Universal 100 to 240 Vac ±10 percent, 50 to 60 Hz, no switches or jumpers
required, ¾-14 national pipe thread (NPT) conduit port
1.3.20 
Pollution degree
2
1.3.21 
Over voltage category
II
1.3.22 
Relative humidity
5 to 95 percent (non-condensing)
1.3.23 
Isolated output
Oxygen
4-20 mAdc, 950 ohm maximum with HART
only
Combustibles
4-20 mAdc, 950 ohm maximum
1.3.24 
Alarm
Alarm output relay - dry contact, form C, 30 mA, 30 Vdc capability
1.3.25 
Power consumption
750 W maximum
1.4 

Ordering information

For ordering information, refer to the
Combustibles Transmitter Product Data
Rosemount OCX8800A
Description and specifications
or F
Fieldbus capability
®
OUNDATION
Rosemount OCX8800 Oxygen and
Sheet.
December 2022