Emerson Rosemount OCX8800A Manual page 23

Oxygen and combustibles transmitter
Table of Contents

Advertisement

Manual
00809-0500-4880
-40 to 185 °F (-40 to 85 °C), internal operating temperature of electronics inside
housing, as read by HART
Local operator interface
-40 to 158 °F (-40 to 70 °C), ambient
At temperatures above 158 °F (70 °C) inside instrument housing, the infrared
keypad will cease to function, but the Rosemount OCX8800 will continue to
operate properly.
1.3.6 
Nominal and approximate shipping weights
18-in. (457 mm) probe package
54 lb. (20 kg)
3-ft (0.91 m) probe package
55 lb. (20.5 kg)
6-ft (1.83 m) probe package
57 lb. (21 kg)
9-ft (2.74 m) probe package
59 lb. (22 kg)
1.3.7 
Housings mounting
Integral electronics
Remote electronics
1.3.8 
Materials
Probes
316L stainless steel: 1300 °F (704 °C)
Inconel 600: 1832 °F (1000 °C)
Ceramic: 2600 °F (1427 °C)
Enclosures
Low copper aluminum
Rosemount OCX8800A
or F
Fieldbus.
®
OUNDATION
Flange
Sensors housing: flange
Electronics housing: wall/pipe
Description and specifications
December 2022
23

Advertisement

Table of Contents
loading