8.7 Configuring basic measurement parameters

The following section describes the settings common for several measurement modes.

8.7.1 Configuring auto averaging

Describes the commands for automatic averaging in continuous average and burst
average measurements.
Remote commands:
SENSe<Sensor>:AVERage:COUNt
SENSe<Sensor>:AVERage:COUNt:AUTO
SENSe<Sensor>:AVERage:COUNt:AUTO:MTIMe
SENSe<Sensor>:AVERage:COUNt:AUTO:NSRatio
SENSe<Sensor>:AVERage:COUNt:AUTO:RESolution
SENSe<Sensor>:AVERage:COUNt:AUTO:SLOT
SENSe<Sensor>:AVERage:RESet
SENSe<Sensor>:AVERage:COUNt:AUTO:TYPE
SENSe<Sensor>:AVERage:TCONtrol
SENSe<Sensor>:AVERage[:STATe]
SENSe<Sensor>:AVERage:COUNt <count>
Sets the number of readings that are averaged for one measured value. The higher the
count, the lower the noise, and the longer it takes to obtain a measured value.
Average count is often also called averaging factor, but it designates the same thing,
i.e the number of measured values that have to be averaged for forming the measure-
ment result.
Averaging is only effective, if
Parameters:
<count>
Example:
SENSe<Sensor>:AVERage:COUNt:AUTO <state>
Sets how the average count is determined.
Parameters:
<state>
User Manual 1179.5760.02 ─ 02
Configuring basic measurement parameters
................................................................................... 74
..........................................................................74
................................................................ 76
.................................................................................... 76
................................................................ 76
................................................................................ 77
...................................................................................77
SENSe<Sensor>:AVERage[:STATe]
Range:
1 to 1048576
*RST:
1024
AVER:COUN 1
ON
Auto averaging: the averaging factor is continuously determined
and set depending on the power level and other parameters.
OFF
Fixed filter: the previous, automatically determined averaging
factor is used.
Remote control commands
...............................................................75
.............................................................75
........................................................ 75
ON is set.