R&S NRP P Series User Manual page 31

Pulse power sensors
Table of Contents

Advertisement

®
R&S
NRPxxP
Example:
You want to update your R&S NRP40P with the NRPxP_02.30.21062301.rsu file.
This file has a size of 10242884 bytes.
To send the file to the power sensor for updating the firmware, your application has to
assemble a memory block containing:
SYST:FWUP <block_data>
The <block_data> is definite length-arbitrary block data as described in
FWUPdate
The size of the file is 10242884. This number has 8 digits. Thus, the <block_data>
consist of the following:
#
8
How many digits follow to specify the file size.
10242884
Number that specifies the file size.
<file_contents>
Contents of the *.rsu file, byte-by-byte
0x0a
Delimiter
In this example, you write exactly 10242905 bytes to the power sensor, for example by
using a 'viWrite()' function.
The 10242905 bytes result from the values of the list above:
9 + 1 + 1 + 1 + 8 + 10242884 + 1
In a (pseudo) string notation, the memory block looks as follows:
SYST:FWUP #810242884<file_contents>0x0a,
User Manual 1179.5760.02 ─ 02
on page 112.
Firmware update
Updating the firmware
SYSTem:
31