Trigger:count; Trigger:delay; Trigger:dtime - R&S NRP P Series User Manual

Pulse power sensors
Table of Contents

Advertisement

®
R&S
NRPxxP
The delay time is set using TRIGger:ATRigger:DELay.
Parameters:
<state>
TRIGger:COUNt <count>
Sets the number of measurement cycles to be performed when the measurement is
started using INITiate[:IMMediate].
This number equals the number of results that can be obtained from the power sensor
after a single measurement. As long as the defined number of measurements is not
executed, the power sensor automatically initiates another measurement internally
when the current result is available.
This command is particularly useful in conjunction with buffered measurements. For
example, to fill a buffer with a predefined size with measurements that have been trig-
gered externally or by
Parameters:
<count>
Example:
TRIGger:DELay <delay>
Sets the delay between the trigger event and the beginning of the actual measurement
(integration).
Parameters:
<delay>
TRIGger:DTIMe <dropout_time>
Sets the dropout time for the internal trigger source. During this time, the signal power
must exceed (negative trigger slope) or undercut (positive trigger slope) the level
defined by the trigger level and trigger hysteresis. At least, this time must elapse before
triggering can occur again.
See
Parameters:
<dropout_time>
User Manual 1179.5760.02 ─ 02
*RST:
*TRG
Range:
*RST:
See
measurement",
Range:
*RST:
Default unit: s
Chapter 8.5.2.4, "Dropout
Range:
*RST:
Default unit: s
OFF
without having to start the measurement multiple times.
1 to 2147483646
1
Chapter 9.2, "Performing a buffered continuous average
on page 120.
-5.0 to 10.0
0.0
time",
on page 44.
0.00 to 10.00
25.0e-9
Remote control commands
Controlling the measurement
51

Advertisement

Table of Contents
loading

This manual is also suitable for:

Nrp18pNrp40pNrp50p1444.1190k021444.1290k021444.1390k02

Table of Contents