NRPxxP
Setup
1
Figure 5-2: Setup with the R&S Power Viewer
1 = Signal source
2 = R&S NRPxxP power sensor
3 = Host interface connector
4 = R&S NRP‑ZKU cable
5 = USB connector
6 = Computer with installed R&S Power Viewer
1. NOTICE! Incorrectly connecting or disconnecting the power sensor can damage
the power sensor or lead to erroneous results. Ensure that you connect or discon-
nect the power sensor as described in
on page 12.
Connect the power sensor to the signal source.
2. Connect the cables as shown in
For a detailed description, refer to
on page 14.
Starting a measurement
For a detailed description, refer to the operating manual of the R&S Power Viewer. The
manual is installed automatically during the installation of the R&S Power Viewer.
1. Start the R&S Power Viewer.
2. Execute zeroing.
Note: Turn off all measurement power signals before zeroing. An active measure-
ment signal during zeroing causes an error.
3. Switch on the test signal of the signal source.
4. Select a measurement.
5. Start the measurement.