R&S
NRPxxP
5. If needed, load uncertainty data. See
on page 90.
6. Check the entries in the " S-Parameter Device Mnemonic", "Lower Power Limit/W"
and "Upper Power Limit/W" fields
7. If necessary, change the entries.
For example, the lower and upper power limits are deduced from the power limits
of the power sensor itself and the minimum attenuation of the S-parameter device.
If the "Upper Power Limit/W" entry is higher than the power dissipation rating of the
attenuator, reduce it accordingly.
8. Click "OK" to apply the changes.
If you want to discard the change, click "Cancel".
To load an uncertainty parameter file
1. In the device table, double-click an entry. See also
2. Select "File" > "Import uncertainties".
3. Select the file you want to import.
4. Confirm with "Open".
The data from the selected file is loaded in the device table.
User Manual 1179.5760.02 ─ 02
Remote control commands
Configuring basic measurement parameters
"To load an uncertainty parameter file"
"Device table"
on page 87.
90