Status Reporting System; Overview - R&S NRP P Series User Manual

Pulse power sensors
Table of Contents

Advertisement

®
R&S
NRPxxP
USB Resource String
The syntax of the used USB resource string is:
USB::<vendor ID>::<product ID>::<serial number>[::INSTR]
where:
<vendor ID> is the vendor ID for Rohde & Schwarz (0x0AAD)
<product ID> is the product ID for the Rohde & Schwarz power sensor
<serial number> is the individual serial number on the name plate at the rear side
of the power sensor
Table 10-2: R&S NRPxxP USB product IDs
R&S NRPxxP power sensor
R&S NRP18P
R&S NRP40P
R&S NRP50P
Example:
USB::0x0AAD::0x0143::100001
0x0AAD is the vendor ID for Rohde & Schwarz.
0x0143 is the product ID for the R&S NRP18P power sensor.
100001 is the serial number of the particular power sensor.

10.2 Status reporting system

The status reporting system stores all information on the current operating state of the
sensor, and on errors which have occurred. This information is stored in the status reg-
isters and in the error queue. You can query both with the commands of the STATus
subsystem.

10.2.1 Overview

Fig.10-1
User Manual 1179.5760.02 ─ 02
shows the hierarchical structure of information in the status registers.
Remote control basics
Status reporting system
USB product ID
0x0143
0x017C
0x017E
126

Advertisement

Table of Contents
loading

This manual is also suitable for:

Nrp18pNrp40pNrp50p1444.1190k021444.1290k021444.1390k02

Table of Contents