Trigger:level; Trigger:level:unit; Trigger:sender:port - R&S NRP P Series User Manual

Pulse power sensors
Table of Contents

Advertisement

®
R&S
NRPxxP
Usage:
TRIGger:LEVel <level>
Effective only if
Sets the trigger threshold for internal triggering derived from the test signal.
If an S-parameter device and/or the offset correction are enabled, the trigger threshold
is referenced to the correction data.
If the S-parameter device and/or the offset correction are disabled, the trigger thresh-
old and its input limits are adjusted as necessary.
If you enter a value without unit, the unit is defined by TRIGger:LEVel:UNIT.
Parameters:
<level>
TRIGger:LEVel:UNIT <unit>
Sets the unit of the trigger level if this value is entered without a unit. See also

TRIGger:LEVel

Parameters:
<unit>
TRIGger:SENDer:PORT <sender_port>
Selects the port where the power sensor outputs its own trigger event in case it is trig-
ger sender. See
If the power sensor is the trigger sender, it can output its trigger event either on the
EXTernal<1> or EXTernal2.
If the power sensor triggers itself, the trigger source of the power sensor must be
assigned to the other external port, as shown in the examples.
Parameters:
<sender_port>
Example:
Example:
User Manual 1179.5760.02 ─ 02
Event
TRIGger:SOURce
INTernal.
Range:
10.0e-7 to 100.0e-3
*RST:
1.0e-4
Default unit: Watts
on page 53.
DBM | W | DBUV
*RST:
W
TRIGger:SENDer:STATe
EXT1 | EXTernal1 | EXT2 | EXTernal2
*RST:
EXT1
TRIG:SEND:PORT EXT1
TRIG:SOUR EXT2
TRIG:SEND:STAT ON
TRIG:SEND:PORT EXT2
TRIG:SOUR EXT1
TRIG:SEND:STAT ON
Remote control commands
Controlling the measurement
for more information.
53