Burst Average Measurement Results; Timeslot Measurement Results; Trace Measurement Results - R&S NRP P Series User Manual

Pulse power sensors
Table of Contents

Advertisement

®
R&S
NRPxxP
Usage:
SENSe<Sensor>:[POWer:][AVG:]BUFFer:DATA?
Queries the results of the continuous average result buffer and returns them even if the
buffer is not full.
In contrast,
the buffer is full.
Usage:

8.9.2 Burst average measurement results

FETCh<Sensor>[:SCALar][:POWer]:BURSt?
Queries the last valid measurement result.
To configure the measurand, use
ated.
Usage:

8.9.3 Timeslot measurement results

SENSe<Sensor>:TRACe:DATA?
FETCh<Sensor>[:SCALar][:POWer]:TSLot?
Queries the last valid measurement result.
To configure the measurand, use
ated.
Usage:

8.9.4 Trace measurement results

SENSe<Sensor>:TRACe:DATA?
SENSe<Sensor>:TRACe:FFT:DATA?
SENSe<Sensor>:TRACe:DATA?
Returns the measured trace data in a well-defined format, showing the course of power
over a defined time.
Unlike FETCh<Sensor>[:SCALar][:POWer][:AVG]?, this command takes the set-
tings of
Command response
User Manual 1179.5760.02 ─ 02
Query only
FETCh<Sensor>[:SCALar][:POWer][:AVG]?
Query only
Query only
Query only
....................................................................................... 96
SENSe<Sensor>:AUXiliary
Querying measurement results
before the measurement is initi-
CALCulate:FEED
on page 96
before the measurement is initi-
CALCulate:FEED
.................................................................................98
into account, as explained below.
Remote control commands
returns a result only if
96

Advertisement

Table of Contents
loading

This manual is also suitable for:

Nrp18pNrp40pNrp50p1444.1190k021444.1290k021444.1390k02

Table of Contents