11.4 Deactivating the Accumulator
To avoid damaging the bed and the environment during storage:
►
Deactivate the accumulator on the Attendant Control Panel.
To deactivate the accumulator on the Attendant Control Panel:
►
Disconnect the bed from the mains.
►
Disconnect the ground wire.
►
Activate the keypad by pressing the GO button on the Attendant Control Panel.
►
Press the Thighrest Up + Thighrest Down + Trendelenburg Position buttons at the same time and hold
them for three seconds.
►
The accumulator is deactivated.
NOTE: Try some functions to ensure the accumulator is deactivated.
To activate the accumulator again:
►
Connect Power Cable to the mains.
12 Manipulation
WARNING!
Risk of injury when adjusting the bed!
►
Ensure there are no body parts between the mattress platform elements and the mattress platform frame when
adjusting the bed.
►
Ensure there are no body parts below the mattress platform frame before adjusting the bed.
►
Secure or remove any items on the bed.
The bed is operated by different control elements.
Control elements:
►
ACP Attendant Control Panel
►
Handset
Disabling individual functions on the Attendant Control Panel affects all of the control elements.
If the bed does not react to individual position settings:
►
Check whether the function is disabled on the Attendant Control Panel.
D9U001AM2-0101_06
33