WARNING!
An additional MULTIPLE SOCKET-OUTLET or extension cord shall not be
connected to the MEDICAL ELECTRICAL SYSTEM.
WARNING!
During specific investigations or treatments the significant risks of reciprocal interference posed by ME equip-
ment may occur.
DANGER!
Risk of injury or death due to use of incorrect equipment!
► Always conduct the risk assessments required for the selection of suitable equipment.
WARNING!
Any serious incident that has occurred in relation to the device should be reported to the manufacturer and the
competent authority of the Member State in which the user and/or patient is established!
WARNING!
Only authorised and trained person using the tool is allowed to change fuses and power supplies!
WARNING!
This medical device is not intended for oxygen enriched environment!
WARNING!
This medical device is not intended for use with flammable substances!
WARNING!
This medical device is not portable medical electrical equipment!
WARNING!
Make sure the duty cycle (2 min ON/18 min OFF) is not exceeded during bed positioning!
WARNING!
Patient is allowed to use selected control elements only if hospital personnel had assessed that the patient´s
physical and psychological state is in accordance with use of them and only if the hospital personnel had
trained the patient in accordance with the instructions for use!