10 Power Supply Cord (Mains Power Cable)
CAUTION!
Disconnecting bed from the mains does not stop motions of the bed!
►
Stop the bed before disconnection bed from the mains.
Attachment plug is means of connecting and disconnecting bed from the mains.
Power supply cable (mains power cable) must be attached with a hook at the head end of the bed during transport.
Where the integrity of the external protective conductor in the installation or its arrangement is in doubt
►
operate the bed from internal accumulator only.
11 Accumulator
CAUTION!
Risk of reducing accumulator durability due to incorrect use!
►
Use bed on accumulator only in crisis situations (e.g.: power blackout, patient complications during transport,
etc.).
►
After reconnecting bed to the mains charge accumulator to full capacity (see chart Accumulator charge status).
CAUTION!
Risk of damage or destruction of accumulator!
►
If the accumulator is faulty, degassing may occur. In rare cases this might cause deformations of the accumulator
case, control panel housing or cable.
►
If this occurs stop using the bed immediately (see Removing the Bed from Service).
►
Inform the manufacturer's service department immediately.
For declared lifetime period of leaded accumulators is recommended during storage:
1.
To prevent accumulators from deep discharging and to keep accumulators at least partly charged by regular recharging
2.
To store accumulators on the dry places with temperature from 10°C to 40°C
3.
To prevent accumulators from being in the sunshine
Accumulator lifetime could be up to 5 years if operated under optimum conditions.
Accumulator capacity can be significantly reduced if:
►
too high ambient temperature
►
many accumulator charge/discharge cycles
►
recurrence of deep discharge
►
bed is often powered only by the accumulator
The accumulator supplied with the bed is delivered uncharged. The accumulator serves as a backup during power failures or while
transporting the patient.
►
Use only accumulators approved by the manufacturer.
►
The manufacturer provides a 6-month warranty for the full function of accumulators.
►
Check the accumulator functionality at least once a month in accordance with the user and service manuals
and have the accumulator changed if necessary.
►
The manufacturer will assume no responsibility for any damage to the bed or the accumulator caused by:
■
Non-observance of the manufacturer's instructions in the instructions for use.
■
Using accumulators not approved by the manufacturer.
■
Accumulator replacement non-qualified service organisation.
To charge the accumulator:
►
Connect the bed to the mains and check the yellow LED on the Attendant Control Panel according to the following table.
NOTE: Some bed adjustment options are not available without a accumulator, for example, height adjustment under a load of
above 200 kg.
►
Charging and accumulator capacity is indicated by the yellow LED placed on the ACP control panel.
D9U001AM2-0101_06
31