Operating Instructions Cubis MSE

Basic Weighing Function

Features
Taring the balance
Printing weights
Using Verified Balances as Legal Measuring Instruments in the EU*:
The type-approval certificate for verification applies only to non-automatic
weighing instruments.
For automatic operation with or without auxiliary measuring devices, you must
comply with the regulations applicable to the place of installation.
§ Before using the balance as a legal measuring instrument, calibrate and adjust
it at the place of use using the built-in motorized calibration weight:
for details, see "Calibration|Adjustment" in this chapter.
$ The temperature range (°C) indicated on the verification label may not be
exceeded during operation.
Example:
MSX
K
+10/+30°C
+5°C/+40°C isoCAL
Preparation
§ Switch on the balance: Press A
§ Level the balance if required
§ Tare the balance, if necessary: Press J
$ If necessary, change the configuration settings: See the chapter entitled "Con-
figuration"
$ If desired, load the factory settings: See the chapter entitled "Configuration"
Additional functions
$ Switching off the balance: Press A
Example: Simple Weighing
Step
1. Switch on the balance
Self-test is performed,
followed by automatic initial
tare function.
2. Place container on weighing pan
(in this example: 11.5 g).
3. Tare the balance
4. Place sample in container
(in this example: 132 g).
5. Print weight.
Key (or Instruction)
Display|Printout
0.0 g
A
+
11.5 g
0.0 g
J
+
132.0 g
N
+
P
132.0 g