Operating Instructions Cubis MSE
Connecting the Device to AC Power
Insert the AC adapter plug into the wall outlet.
t
y The balance is now ready to use.

Warm-up Time

y To return precise results, the device must warm up for at least 30 minutes
after initial connection to the power supply.
Only after this time will the device have reached the required operating temper-
ature.
The first time a verified balance used in legal metrology (legal-for-trade
applications) is connected to the power, it must warm up for at least
24 hours before operation.
After having been disconnected and then connected again, it must
warm up for 30 minutes.