Sartorius Cubis Series Operating Instructions Manual page 48

Electronic precision and analytical balances
Hide thumbs Also See for Cubis Series:
Table of Contents

Advertisement

Using the Balance
Example:
Calculating the weight per unit area of paper: An A4 sheet of paper is used in this example, with surface dimensions of 0.210 m
+ 0.297 m = 0.06237 m
2
. To determine the weight per unit area, the total weight is divided by the surface.
Parameter settings:
Applic. - Calc. - method - Divis.
Step
1.
Tare the balance
2.
Activate divisor input
3.
Set the divisor (in this example: 0.06237):
Position the decimal point,
Enter numerals in increments of 1 or:
4.
Store the divisor and initialize the balance.
The current divisor remains stored in battery-backed
memory until the setting is changed.
5.
Weight per unit area: Place an A4 sheet of paper on
the balance
6.
If desired, print result
7.
Toggle display between weight and calculated
value
8.
Unload the balance
9.
Repeat as needed, starting from Step 5
48
Operating Instructions Cubis MSE
Key (or Instruction)
J
S
V, 5+ S,
2+ V,
Repeatedly press S
briefly or hold the S
key
V
P
Repeatedly S
Display|Printout
0.00 g
______1.0
___.00000
___.06000
___.06237
+
o
0.0
Div
0.6237
o
+
79.7
Res
+
79.7 o
+
4.97 g
+
o
79.7
+
o
0.0

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents