Power Connection

Check the voltage rating and plug design!
t
The printed voltage rating (see type label) must match the voltage in the place
of installation.
If the stated supply voltage or the plug design of the power cord does not
comply with your country's standard, please inform the nearest Sartorius repre-
sentative or your dealer.
The power connection must be made in accordance with the regulations appli-
cable in your country.
In order to connect the equipment to the power supply (protection class 1),
use a correctly installed wall outlet with a protective grounding conductor (PE)
and a fuse of a maximum 16 A.
If a main feeder from the ceiling is required or if installation of a CEE plug is
necessary, this must be carried out by a qualified electrician.
Only use original Sartorius power cables!
Safety Precautions
The output line of the power supply has a pole (GND) connected to the metal
balance housing. The data interface is also galvanically linked to the balance
housing (GND).
1. Plug the DC supply cable into the scale and tighten the threaded fitting.
1
2. Plug the other end of the power cord into the AC adapter.
2
Getting Started
Operating Instructions Cubis MSE
17