Installing The Ac Option Power Supply - Philips CARDIOMD Service Manual

Table of Contents

Advertisement

12.6.2.2 Installing the AC Option Power Supply

Procedure
1. Remove the back cover from the table
console.
2. On the system's main power supply
(PSU), disconnect the orange 3-way plug
from PSU connector C6 (24V EM-ST).
3. On the main PSU, disconnect the white 3-
way plug from PSU connector OC1
(115VAC) and unscrew the associated
ground wire.
4. Place the AC Option power supply
(9POW1315) on the floor next to the main
PSU.
5. Now plug the two connectors that you just
removed from the main PSU into the
mating headers on the AC Option power
supply.
6. Attach the ground wire to the available
post.
Note. There are two of each type of header.
Headers of identical type can be swapped as
they are connected in parallel.
1MAN0158-H11 04-2005 CSIP Level 1