Configuring Antivirus Software - Philips CARDIOMD Service Manual

Table of Contents

Advertisement

Acquisition PC Software Updates
Procedure
communication is working properly. In-
structions are provided in Chapter 5 Diag-
nostics.
Installing the CardioMD Application Software
Procedure
1. Insert the CardioMD software CD in the
acquisition PC's CD-ROM drive.
2. Close all programs on the acquisition PC.
3. Start the Windows Explorer and select the
CD-ROM drive.
4. Run the file
InstallCardioMDto-
by double-clicking on the name
tal.bat
of the file.
The file is located in the root directory of the
CD-ROM.
5. Follow the instructions given.
6. When the software installation is com-
pleted, restore calibration tables and cus-
tomer specific protocols from the backup
disks. Instructions are provided in the Ap-
pendix of the CardioMD Operator's Man-
ual.
7. Finally, remove the software CD-ROM
and store it in a safe place along with the
backup disks.
8. Start the CardioMD acquisition software
by double-clicking the CardioMD icon on
the acquisition PC's desktop.
7.3.3

Configuring Antivirus Software

Acquisition PC's running Windows XP (CardioMD Series III systems) are supplied with antivirus
III
software. This section provides instructions for configuring Symantec AntiVirus™ on the acquisi-
tion PC.
If the CardioMD acquisition PC has no access to the internet or outside networks, you may want
to prevent it from trying to automatically check for new antivirus updates through its LiveUpdate
feature. To disable LiveUpdate, follow the instructions below.
7-22 CardioMD Service Manual
Details
Details
1MAN0158-H11 04-2005 CSIP Level 1

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents