Detailed Description
Attenuator in closed position
Figure 12.20 Attenuator
Preparing for Transportation
When the line source housing is going to be transported, e.g. being shipped from the supplier to
the customer site, the shutter is locked in the closed position. This prevents uncontrolled radiation
from the source if the shutter lever on the line source housing should be activated by mistake.
The shutter is locked by inserting two 3 mm transportation screws into the line source housing and
through dedicated holes in the shutter plate as illustrated in Figure 12.21.
Figure 12.21 Locking the shutter for transportation
Covers
The AC option includes two aluminum covers, both mounted on the line source compartment:
A small edge cover
A larger service cover.
These are shown in Figure 12.12, page 12-16.

12.3.4.2 Line Source Compartment Electronics

The printed circuit board inside the line source compartment, the AC source motion board, carries:
The line source motion controller
A driver circuit for the solenoid
An interface to the collision pad.
The board has a DB15 connector for the flexible cable to the AC bracket.
12-22 CardioMD Service Manual
Transportation screw
Attenuator retracted
Locking hole
1MAN0158-H11 04-2005 CSIP Level 1