8.1 Installation Requirements

8.1.1

Introduction

This installation section describes the installation of CardioMD Series III system. Should you
I, II
III
need to install a CardioMD Series I or II system, refer to Chapter 8 Installation of the service
manual originally provided with the CardioMD Series I or II system.
Note. A few of the photos in this chapter actually show CardioMD Series I/II systems. This is the
case only where these details of the system are insignificant for the installation procedure.
Note. Regardless of the system version, read the document CardioMD Camera Site Planning, part
no. 9347-0112.
Note. Section 8.3.4 Software Setup on page 8-75 does provide some information on setting up
I, II
software under Windows 2000, which is used on Series I and II systems acquisition PC's.
8.1.2

Ideal Floor Space Allocation

Some general considerations should be followed for all nuclear camera systems. These include
comfortable workspace, good visual access to patients on the equipment, sufficient service access,
and patient safety. All nuclear equipment contains components such as the sodium iodide crystals
and high speed electronics that require reasonably well controlled temperature, humidity, and
power.
Complete camera systems can be installed in a single room or with the gantry and acquisition
station in one room and the processing station in a separate room.
8.1.3

Minimum Floor Space Allocation

Minimum floor space layouts are shown for the CardioMD system including processing station
electronics.
The electronics do not require a separate environmentally controlled room, provided that the
proper cooling and humidity ranges are available.
#
Equipment
1a
Gantry and table assembly*
Incl. +/– 5 cm travel
1b
Gantry and table assembly*
Incl. +/– 5 cm travel
2
Collimator cabinet
3a
Cart-mounted acquisition PC
3b
Wall-mounted acquisition PC
3c
Desk-mounted acquisition PC
*
The CardioMD system is available in a standard configuration (1a) and a reduced room size version (1b)
Table 8.1 Minimum floor space allocation for the CardioMD system
1MAN0158-H11 04-2005 CSIP Level 1
Width
Depth
112 inch
58 inch
284 cm
150 cm
83 inch
58 inch
210 cm
150 cm
17 inch
17 inch
43 cm
43 cm
25.2 inch
25.4 inch
63.9 cm
64.5 cm
30.8 inch
15.2 inch
79 cm
39 cm
22.8 inch
15.0 inch
58 cm
38 cm
Installation
Height
Weight
58 inch
1797 lbs.
150 cm
815 kg
58 inch
1797 lbs.
150 cm
815 kg
40 inch
75 lbs.
102 cm
33 kg
47.3 inch
66 lbs.
120 cm
30 kg
22.6 inch
51 lbs.
58 cm
23 kg
11.8 inch
10 lbs.
30 cm
5 kg
CardioMD Service Manual 8-3