Environmental Specifications; Networking And Cables; Networking - Philips CARDIOMD Service Manual

Table of Contents

Advertisement

8.2.5

Environmental Specifications

The temperature in the examination room must be regulated at 59° F to 86° F (15° C to 30° C).
The humidity must be regulated at 45 to 80 % relative humidity (non-condensing). 55 % relative
humidity is optimum. The tables below indicate the heat dissipation of the components of the
CardioMD system.
CardioMD system
2,360 BTU/hr
165 Cal/sec
691 watts
Table 8.8 Heat dissipation of CardioMD system components
2.2 kVA (APC Inc., optional)
275 BTU/hr
19 Cal/sec
80 watts
Table 8.9 Heat loads for the UPS options
8.2.6

Networking and Cables

8.2.6.1

Networking

There are two general types of CardioMD networking installations:
1. Single camera/computer
2. Multiple camera/computer.
Single Camera/Computer
In a single camera/computer installation, Philips Medical Systems will network together the
CardioMD system, the PegBlade computer and a printer.
The customer is responsible for providing one twin RJ-45 connection box near the CardioMD
system and one near the PegBlade. The customer also needs to run Cat 5 cables in the walls,
ceilings, or floors between the connection boxes.
If the customer does not provide the connection boxes and does not run the Cat 5 cables in the
walls, Philips will run those cables along the floor. That may be unsightly if the connection boxes
are far apart.
Multiple Camera/Computer
In a multiple camera/computer installation, Philips Medical Systems will network together the
CardioMD system, the PegBlade computer and perhaps a printer and other networked devices.
Other networked devices could be other CardioMD cameras, other Philips cameras, a hospital
network, a network printer or a network viewing station as shown in the examples in Figure 8.10
on page 8-18.
1MAN0158-H11 04-2005 CSIP Level 1
Jetstream Cardio (optional)
CPU
Monitor
1,400 BTU/hr
136 BTU/hr
98 Cal/sec
10 Cal/sec
410 watts
40 watts
3.1 kVA (Best Inc., optional)
742 BTU/hr
52 Cal/sec
217 watts
PegBlade (optional)
CPU
Monitor
854 BTU/hr
544 BTU/hr
60 Cal/sec
38 Cal/sec
250 watts
159 watts
5.0 kVA (Toshiba, optional)
2,610 BTU/hr
183 Cal/sec
765 watts
CardioMD Service Manual 8-17
Installation